Features

  • Certify IR pyrometers from –30 °C to 500 °C (–22 °F to 932 °F)
  • Large 57 mm (2.25 in) blackbody target
  • RTD reference well for high precision
  • Small, compact design
Specifications 9132

Temperature Range

50 °C to 500 °C
(122°F to 932°F)

Accuracy 
(does not include emissivity)

±0.5 °C at 100 °C
(±0.9°F at 212°F) 
±0.8 °C at 500 °C
(±1.4 °F at 932 °F)

Stability

±0.1°C at 100°C
(±0.18°F at 212°F) 
±0.3°C at 500°C
(±0.54°F at 932°F)

Target Size

57 mm (2.25 in)

Target Emissivity

0.95 (±0.02 from 8 to 14 mm)

Resolution

0.1 °

Heating Time

30 minutes (50 °C to 500 °C)

Cooling Time

30 minutes (500 °C to 100 °C)

Computer Interface

RS-232 I/O included with 9930 Interface-it software

Power

115 VAC (±10 %), 3 A or
230 VAC (±10 %), 1.5 A, switchable, 50/60 Hz, 340 W

Size (HxWxD)

102 x 152 x 178 mm
(4 x 6  x 7 in )

Weight

1.8 kg (4 lb)

NIST-Traceable Contact Calibration

Data at 50°C, 100°C, 200°C, 250°C, 300°C, 400°C, and 500°C

Whether you’re using in-line or handheld infrared pyrometers, you need good calibration standards to verify their accuracy. Our portable IR calibrators provide stable blackbody targets for calibrating noncontact IR thermometers from –30°C to 500°C.

These units feature a temperature controlled measurement surface with a diameter of 2.25" (57 mm). The target temperature can be controlled in set-point increments of 0.1° from –30°C to 500°C and a well is located directly behind the blackbody surface for contact calibration.   9132 with IR

9132

For IR calibrations above normal ambient, the 9132 provides a stable measurement surface up to 500°C (932°F). Short heating and cooling times mean you won’t have to wait long to get your work done.