Fluke 5609-15-A Probe, Secon. PRT



 

Features

  • 5609: –200 °C to 670 °C (100 mm minimum immersion)
  • ± 0.01 short-term stability
  • Calibration not included, NVLAP-accredited calibration
    optional, lab code 200348-0
  • Dry-well reference with 90 ° bend

Specifications

Temperature range

5609: –200 °C to 670 °C

Nominal resistance at 0.01 °C

100 W ± 0.5 W

Temperature coefficient

0.0039250 W/W/°C

Accuracy[1]

See footnote

Short-term repeatability[2]

± 0.01 °C at 0.010 °C
± 0.02 °C at max temp

Drift[3]

± 0.01 °C at 0.010 °C
± 0.02 °C at max temp

Hysteresis

± 0.01 °C maximum

Sensor length

30 mm ± 5 mm (1.2 in ± 0.2 in)

Sensor location

3 mm ± 1 mm from tip (0.1 in ± 0.1 in)

Sheath material

Inconel™ 600

Minimum insulation resistance

5609: 500 MW at 23 °C, 10 MW at 670 °C

Transition junction
dimensions

71 mm x 12.5 mm (2.8 in x 0.49 in)

Minimum immersion length[4] (<5 mK error)

5609: 100 mm (3.9 in)

Maximum immersion length

305 mm (12 in)

Response time[4]

5609: 12 seconds typical

Self heating (in 0 °C bath)

5609: 50 mW/°C

Lead-wire cable type

Teflon,™ 24 AWG

Lead-wire length

1.8 m (6 ft)

Lead-wire temperature range

–50 °C to 250 °C

Calibration

Calibration not included; NVLAP-accredited calibration optional, lab code 200348-0. Please see calibration uncertainty table and its explanation of changeable uncertainties.

If you want a very stable thermometer from –200 °C to 670 °C and are particular about the calibration, then look no further than our 5609 Secondary Reference PRT. Its short-term stability at the triple point of water is only ± 7 mK and its one-quarter inch (6.23mm) diameter lets you get accurate measurements in only 100 mm of immersion.

The 5609 comes with a one-quarter inch (6.35 mm) diameter sheath in lengths of 12 inches, 15 inches, and 20 inches; or with a 6 mm diameter in lengths of 300 mm, 400 mm, or 500 mm.

When looking for improved response time and reduced stem effect in shallow immersion, look for small diameter probes, because the measurement error called stem effect is caused by the diameter of the stem rather than the length of the stem.

These probes have Inconel™ sheaths and are made using a special manufacturing process, giving them great precision over a wide temperature range. The sensors for these probes are reference-grade platinum and feature four-wire connections with less noisy measurements than two-wire counterparts.

These probes come with a certificate of compliance to ensure the meet their specifications. If you’d like, you can also order a NVLAP-accredited calibration from our laboratory, lab code 200348-0. On the report of calibration, you’ll get the test data and the ITS90 calibration coefficients that you can easily input into any Hart thermometer. If you order your probe with an INFO-CON connector, we’ll program the coefficients directly into your connector, which loads the coefficients for you when you plug it into our 1522 Handheld Thermometer.







You may also be interested in this/these product(s)

Fluke 2609 Long Probe Carrying Case
Fluke 2609 Long Probe Carrying Case
Fluke 5609-15-B Probe, Secon. PRT
Fluke 5609-15-B Probe, Secon. PRT
Fluke 5609-15-D Probe, Secon. PRT
Fluke 5609-15-D Probe, Secon. PRT
Fluke 5609-15-G Probe, Secon. PRT
Fluke 5609-15-G Probe, Secon. PRT

buy xenical | buy xenical | buy zoloft | buy lipitor | buy allegra | buy clomid | buy celebrex | buy amoxicillin